• Part: SGX-7CL2
  • Description: Industrial CL2 Sensor
  • Manufacturer: SGX
  • Size: 308.33 KB
Download SGX-7CL2 Datasheet PDF
SGX
SGX-7CL2
SGX-7CL2 Datasheet Industrial CL2 Sensor PERFORMANCE Output signal Typical Baseline Range (pure air) T90 Response Time Measurement Range Maximum Overload Linearity Repeatability Remended Load Resistor Resolution (Electronics dependent) 1000 ± 300 n A / ppm ±0.2ppm CL2 equivalent < 60 seconds 0 - 20 ppm 250 ppm Linear < ±2% CL2 equivalent 33 ohms 0.1 ppm typical OPERATING CONDITIONS Temperature Range -20°C to +50°C Pressure Range Operating Humidity Range 800 to 1200 mbar 15% to 90% RH LIFETIME Long Term Output Drift Remended Storage Temp Expected Operating Life < 20% per annum 0°C to 20°C > 12 months in air SGX Europe Sp. Z o.o. Building 11 Ligocka St. 103, 40-568 Katowice, Poland T: +48 (0) 32 438 4778 E: sales.is@sgxsensortech. .sgxsensortech. OUTLINE All dimensions are in mm All tolerances are ±0.15mm INTRINSIC SAFETY DATA Max at 150ppm 0.3 m A Max o/c Voltage 1.3 V Max s/c Current <1.0 A © SGX Sensortech 2012-2020 DS-0339 SGX-7CL2, Issue 1, 27-Mar-2020, Page...